99久久国产综合精品麻豆,亚洲中久无码永久在线观看软件,秋霞鲁丝片一区二区三区,亚洲综合天堂一区二区三区

資料下載

您的位置

首頁 資料下載下載中心
石英晶體微天平QCM金晶片的校準
提 供 商: 美國Gamry電化學 資料大小:
圖片類型: 下載次數: 61 次
資料類型: PDF 瀏覽次數: 2636 次
相關產品:

Introduction
This Application Note is the second in a series relating to the eQCM 10M™ and assumes that
you have read and understand the material covered in the Application Note, Basics of a
Quartz Crystal Microbalance.
The purpose of this Application Note is to show the reader a simple experiment where we
reduce Cu2+onto an Au electrode and then oxidize Cu back off. We will then use the result to
calculate a calibration factor for the crystal. Conversely, we will show you how to calculate the
molar mass of the species deposited using a calibration factor.
Each crystal has a theoretical calibration factor, however, under typical experimental
conditions, these calibration factors will vary slightly. We will use Cu2+ to calculate a
calibration factor which could then be used in subsequent experiments on the same crystal.
Cu2+ is reduced in a two electron reaction as shown below:

CuSO4(aq) + 2 e- →Cu(s) + SO42-(aq)
Our goal is to reduce Cu2+ onto an Au electrode, then use the frequency decrease and the
charge passed to calculate a calibration factor.

美國Gamry電化學關鍵詞:多通道電化學工作站,電化學工作站價格,石英晶體微天平,電化學工作站廠家,電化學工作站品牌
版權所有 總訪問量:347310 地址:上海市楊浦區(qū)逸仙路25號同濟晶度310室 郵編:200437
聯系人:客服 郵箱:jqiu@gamry.com
GoogleSitemap 技術支持:化工儀器網 管理登陸 滬ICP備15019588號-2